Preliminary evaluation of a rapid lateral flow calprotectin test for the diagnosis of prosthetic joint infection.
Bone & joint research
According to ICM, 24 patients were defined as PJI positive and the remaining 45 were negative. The overall accuracy of the lateral flow test compared to ICM was 75.36% (52/69, 95% CI 63.51% to 84.95%), sensitivity and specificity were 75.00% (18/24, 95% CI 53.29% to 90.23%) and 75.56% (34/45, 95% CI 60.46% to 87.12%), respectively, positive predictive value (PPV) was 62.07% (18/29, 95% CI 48.23% to 74.19%) and negative predictive value (NPV) was 85.00% (34/40, 95% CI 73.54% to 92.04%), and area under the receiver operating characteristic (ROC) curve (AUC) was 0.78 (95% CI 0.66 to 0.87). Patient data from discordant cases were reviewed by the clinical team to develop the ICM-CR gold standard. The lateral flow test performance improved significantly when compared to ICM-CR, with accuracy increasing to 82.61% (57/69, 95% CI 71.59% to 90.68%), sensitivity increasing to 94.74% (18/19, 95% CI 73.97% to 99.87%), NPV increasing to 97.50% (39/40, 95% CI 85.20% to 99.62%), and AUC increasing to 0.91 (95% CI 0.81 to 0.96). Test performance was better in knees (100.00% accurate (17/17, 95% CI 80.49% to 100.00%)) compared to hips (76.92% accurate (40/52, 95% CI 63.16% to 87.47%)).
Bone & joint research
View Publication